322 HPLC Pump
The 321 and 322 Pumps are multi-solvent pumps designed for use in Gilson’s HPLC analytical to semi-preparative systems. The pumps accommodate flow rates from 0.15 mL/min up to 30 mL/min at pressures up to 8,700 psi (600 bar) using columns up to 21.2 mm ID. A single 321 or 322 pump is necessary for isocratic and binary gradients; tertiary gradients require combining two pumps.
Quick Response
High-pressure Adjustable Volume Dynamic Mixer (AVDM™) enables fast gradient response times and homogeneous mixtures for analytical to semi-preparative applications.
The Right Pump for Your Task
Choose the correct pump head for your application:
• H1 pump heads - ideal for 2-12 mm ID columns with flows up to 15 mL/min and pressures up to 8,700 psi (600 bar)
• H2 pump heads - ideal for 3-20 mm ID columns with flows up to 30 mL/min and pressures up to 4,300 psi (300 bar)
The pump heads are easy to access and simple to maintain, and the 321 and 322 pumps are stackable to conserve bench space.
Easy Control
The 321 functions as a controller with integrated keypad. The 322 is remotely controlled via the 321 controller or TRILUTION® software. Keypad controls include safety files, anomaly processing procedures, and GLP functions. TRILUTION software provides control within a complete prep workstation and includes data processing and results reporting.
Specifications
321 & 322 Pump | ||
Manufacturing Standards | Meet applicable Safety and EMC certification standards; UL and CE certified | |
Environmental Conditions | Indoor use | |
Altitude: up to 2000 m | ||
Temperature range: 5–40°C | ||
Air pressure: 75–105 kPa | ||
Pollution degree: 1 or 2, in accordance with IEC 66 | ||
Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C | ||
Liquid-Contact Materials | 316L stainless steel, titanium, FEP, PTFE, PCTFE, ETFE, PEEK, UHMWPE, sapphire, ruby, and ceramics | |
Power Requirements | Frequency: 50–60 Hz | |
Voltage: 90–120 or 220–260V; mains voltage fluctuations not to exceed ±10% of the nominal voltage | ||
Pressure | Nominal range from 0.5 MPa (5 bar, 70 psi) up to 60 MPa (600 bar, 8,700 psi) for H1, up to 30 MPa for H2 | |
Pump Options | 322-H1*:15 mL/min,60 MPa (8,700 psi) | |
322-H2*:30 mL/min,30 MPa (4,350 psi) | ||
Software | Controlled by TRILUTION® LC Software and 506C interface or GSIOC slave bus | |
Dimensions (w x d x h) | 321: 26 x 41 x 51 cm (10.2 x 16.2 x 20.1 in.) | |
322: 26 x 41 x 38.7 cm (10.2 x 16.2 x 15.3 in.) | ||
Weight | 30 kg (66.1 lbs.) |
*H1 and H2 are interchangeable pump heads. |
Supporting Literature
User's Guides | |
---|---|
321 and 322 Pumps User Guides | Download (EN) |
Technical Product Documentation | |
321 and 322 Pumps Installation Qualifications | Download (EN) |
2023 Gilson Product Guide
The 2023 Gilson Product Guide is available now. Discover best sellers for your Liquid Handling, Purification, and Extraction needs!
DownloadGilson Services and Calibration Programs
Gilson is committed to quality and complete customer satisfaction with all of our products and services. Get system support, training programs and pipette service from the people who know Gilson products the best. We are here to help and provide you with dependable solutions to accommodate your needs.
LEARN MORECOVID-19 Testing
COVID-19 Testing: Liquid Handling Solutions to Support Your Efforts